Illumination-coupling bias breaks rotation equivariance in x-ray-scattering tensor tomography with a circular grating
Phys. Rev. Applied 26, L011005 – Published 24 July, 2026
DOI: https://doi.org/10.1103/tsy5-6mvq
Abstract
X-ray-scattering tensor tomography requires dark-field measurements in the projection domain that are both unbiased and rotation equivariant. We show that, in the context of circular-grating-based tensor tomography, the commonly used cell-based chord retrieval (CCR) cannot satisfy either requirement. From the two-dimensional convolution model of circular gratings, we derive a closed-form expression demonstrating that CCR introduces an illumination-coupling term that directly contaminates the measured dark-field amplitude. This illumination-dependent contribution is fixed in the grating frame and therefore does not co-rotate with the sample, producing orientation-dependent errors and breaking rotation equivariance. In contrast, global two-dimensional Fourier transform retrieval cancels illumination through the convolution theorem and is inherently rotation equivariant. Experiments with a laboratory circular-grating-based tensor tomography system confirm both effects. Illumination coupling is thus an intrinsic limitation of CCR and prevents it from achieving rotation-faithful quantitative anisotropic dark-field imaging.