Radio-Frequency Coulomb-Blockade Thermometry
Phys. Rev. Applied 17, L011003 – Published 26 January, 2022
DOI: https://doi.org/10.1103/PhysRevApplied.17.L011003
Abstract
We present a scheme and demonstrate measurements of a Coulomb-blockade thermometer (CBT) in a microwave-transmission setup. The sensor is embedded in an resonator, where is determined by the conductance of the junction array of the CBT. A transmission measurement yields a signal that is directly proportional to the conductance of the CBT, thus enabling the calibration-free operation of the thermometer. This is verified by measuring an identical sensor simultaneously in the usual dc setup. The important advantage of the rf measurement is its speed: the whole bias dependence of the CBT conductance can now be measured in a time of about 100 ms, which is 1000 times faster than in a standard dc measurement. The achieved noise-equivalent temperature of this rf primary measurement is about at the bath temperature .